Titel
8th Needle/Leaf Interlaboratory Comparison Test 2005/2006
Verfasser
Körperschaft
United Nations Economic Commission for Europe
Forest Foliar Co-ordinating Centre, Vienna
Erscheinungsort
Wien
Verlag
Natural Hazards and Landscape Federal Research and Training Centre for Forests
Erscheinungsjahr
2006
Seiten
33 S. + 74 S. Annex -Results
Material
Monographie
ISBN
3-901347-60-7
Standardsignatur
15015
Datensatznummer
128629
Quelle
Needle/Leaf Interlaboratory Comparison Test
Abstract
The results of the 8th Interlaboratory Comparison Test show generally a good analytical quality in foliar analysis. Some of the laboratories had to learn from their ringtest results, especially those with statistic outliers and/or results outside of the tolerable limits. A few laboratories must change their methods (e.g. dry ashing, nephelometry determination of sulphur) or check their methods with standard reference materials, ring test samples or pirmary standars. Calibration Problems of element analysers appear to be corrected. Most of the labs (90 %) are using control charts as a daily routine procedure. A trend in the use of analytical methods can be seen: - For C,N, (S) element-analyzers are becoming more and more important. - Acid digestion methods in closed systems in combination with ICP methods are very good for the determination of S, P, K, Ca, Mg, Fe, Zn, B and Cu. Flameless-AAS and ICP-MS methods should be used for analysing Cd, Pb and Cu (especially for low concentrations). An X-ray fluorescence analysis is the method of generate good results for S, P, Ca, Mg, K, Zn and Mn.